Draining:
With the switch at "DRAIN", hole 2 is connected to the
drainage hole in the base of the Vapor. The anaesthetic
agent from the Vapor flows through hole 2 and through
the inner tube of the filling adaptor into the bottle.
Saturated vapour from the bottle is routed into the Vapor
through hole 1.
Operation:
The valve is switched to »0« and the sealing block on the
filling device is replaced to seal the three holes 1, 2 and 3
when Vapor is to operate.
1
2
3
0
D
R
A
I
N
41
Description
Key-indexed filling system
1
2
3
0
D
R
A
I
N