Disinfecting
Use surface disinfectants for disinfection. For reasons of
material compatibility use disinfectants based on:
– aldehydes,
– alcohols,
– quaternary ammonium compounds.
Damage to material may occur if preparations are used
which are based on:
– halogen-releasing compounds,
– strong organic acids,
– oxygen-releasing compounds.
We recommend that disinfectants on the current DGHM
(DGHM: German Society for Hygiene and Microbiology)
list are used.
The DGHM list (published by mhp-Verlag GmbH, Wies-
baden, Germany) gives the composition of each disinfec-
tant. For countries where the DGHM list is not familiar,
we recommend the types of disinfectants given above.
Do not immerse Vapor and filling adaptor in liquid and
do not sterilize.
Do not allow disinfectants and cleaning agents to get
into the gas inlet or gas outlet, or the filling system.
Wipe off any residues.
Wipe disinfecting
●
e.g. with Buraton 10 F (Messrs. Schülke & Mayr,
Norderstedt, Germany) or
Cidex (Messrs. Johnson & Johnson, Norderstedt,
Germany).
Follow manufacturer's instructions.
Disinfecting in Aseptor disinfecting machine together
with the anaesthetic machine.
●
Wipe off any dirt with disposable cloth.
●
Vapor must be connected to the anaesthetic machine
and Vapor can remain filled.
●
Turn handwheel to »0« –
»0« button engaged.
●
Disinfect Vapor on anaesthetic machine in
accordance with Aseptor, Instructions for Use.
●
Do not use any programme which has temperatures
over 35
°
C.
23
Care
Disinfecting