When the maximum mark has been reached:
1 Swing bottle down and allow anaesthetic agent to
flow back into the bottle from the adaptor.
Do not fill Vapor above maximum mark.
A shut-off device prevents filling above the maximum
mark. However, if the adaptor is not properly
connected to the bottle or the Vapor, or if the
handwheel is not set at »0«, the anaesthetic agent may
continue to flow into the Vapor.
If Vapor is filled above the maximum mark, the anaes-
thetic agent can flow to the outside via an overflow
drain at the back of the key-indexed filling system.
Then:
●
Swing bottle down.
●
Set fill/drain valve to »DRAIN« and drain anaesthetic
agent into the bottle until maximum mark is reached.
2 Turn fill/drain valve back to »0«.
3 Unscrew locking screw to the stop.
4 Pull adaptor out.
5 Push sealing block in fully.
6 Tighten locking screw.
If this is not done, fresh gas and anaesthetic agent
vapour may escape when Vapor is next switched on.
●
Close anaesthetic agent bottle.
●
Allow any residues of anaesthetic agent in the adaptor
to evaporate under an extractor fan.
3
4
2
0
D
R
A
I
N
Preparation
Filling Vapor
12
6
5
0
D
R
A
I
N
1
0
D
R
A
I
N
max.
دﯾﯾﺎﻣرﻓ ﮫﻌﺟارﻣ.WWW.ELM-FAN.IRتﯾﺎﺳ ﮫﺑ رﺗﺷﯾﺑ دراوﻣ دوﻠﻧاد یارﺑ
To download more items, go to ELM-FAN.IR