Connecting Vapor
Vapor to plug-in system
Use Dräger plug-in system only.
●
Handwheel to »0«, »0« button engaged.
1 There must be undamaged sealing rings on both pins
as fresh gas or anaesthetic agent may escape
otherwise.
2 Locking lever in position over handwheel.
●
Anaesthetic ventilators with two plug-in systems and
Interlock 2:
Move the latch on the Interlock 2 to the opposite
position before hooking in the Vapor.
●
Hold Vapor in vertical position with both hands and
lower gently towards the plug-in system.
The holes of the plug-in adaptor must fit on the pins on
the plug-in system to avoid any damage to the sealing
surfaces.
The Vapor must be level and stable on the plug-in
system, otherwise a loss of fresh gas, leaks or
excessively low output concentrations may result.
If this is not the case:
Disconnect Vapor, as described on page 18, and then
re-connect.
3 Turn locking lever 90
°
clockwise, until it engages.
The Vapor is then secured and cannot be removed.
Vol. %
3
0
Preparation
Connecting Vapor
14
1
0
2
Vol. %
Vol. %
0