Method of Operation
Handwheel
to adjust the concentration between 0.2 (0.3) vol.% and
maximum concentration (depending on type).
The handwheel is locked in the zero setting and can be
freed by pressing »0« button.
Setting concentration:
1 Press »0« button and
2 turn handwheel anti-clockwise to concentration of
anaesthetic agent required.
Plug-in system
to connect and exchange the Vapor quickly and safely.
The Vapor has a permanently fixed plug-in adaptor with
holes which fit over the pins on the plug-in system on the
anaesthetic machine. The adaptor is secured with a
locking lever.
3 The sealing rings become leak-tight when they are
pressed down on the pin by the weight of the Vapor.
Vapor 19.3 and Vapor 19.1 require different plug-in
adaptors.
1
2
5
Method of Operation
Handwheel
Plug-in system
0
Vol. %
3
3