Vapor with filling spout
1 Turn handwheel to »0« – until it engages.
2 Hold an empty bottle for the appropriate anaesthetic
agent under the outlet nozzle of the filling spout.
3 Open outlet valve by about three revolutions – do not
screw valve out.
When no more anaesthetic agent drains from the outlet
nozzle, the Vapor is empty except for the anaesthetic
agent remaining in the wick.
Do not fill bottle right to the top with anaesthetic agent.
Then:
3 Close outlet valve again.
●
Mark bottle: "Used anaesthetic agent".
Storage
See "Technical Data", page 27 for ambient conditions.
●
Set handwheel to »0« –
»0« button is engaged.
●
Drain Vapor, page 19 or 21.
●
For storage periods of more than 6 months, blow off
anaesthetic agent in wick. To do so: set at 4 vol.%
and flush for about 4 hours with 4 L/min Air.
Make sure gas flows into the scavenging system.
●
Before renewed Readiness for Operation, carry out
inspection and service.
21
Operation
Draining Vapor
Storage
3
2
0
1
Vol. %