Technical Manual
|
Dräger Polytron
®
8000 Series
71
Interface settings
The following table shows the correlation between Display indication, relays and
analog interface output depending on assumed values.
Assumed values:
LC sensor (measuring range: 5-10% LEL),
A1 = 8% LEL,
A2 = 9% LEL,
FSD = 7% LEL.
13.1.2 Setting
fault
current
This function defines the current for the fault signal.
1. Select
Settings
>
Communication
>
Analog interface
>
Fault current
and
confirm.
2. Select the line for editing the current and confirm.
3. Set the current and confirm.
The setting for the
Fault current
is displayed.
4. Select
Confirm
and confirm with [OK].
13.1.3
Information regarding the warning signal
To transmit a warning signal via the analog interface, the warning signal must be
switched on.
The warning signal alternates between the warning current and the measurement
current.
– Warning current (interval T2)
– Measurement current (interval T1-T2)
Gas Concentra-
tion
Relays
Display
Analog interface
6.5 % LEL
no change
Measured value
Linear current between 4
and 20mA
7.5 % LEL
no change
Measured value and
20mA
8.5 % LEL
A1 switches
Measured value and
20mA
9.5 % LEL
A1 and A2 switch
Measured value and
20mA
10.5 % LEL
A1 and A2 switch 4 upwards-pointing arrows
Value is over measuring range of
the sensor.
20mA