Troubleshooting
10.2 Warnings
Polytron
®
87x0 IR-specific:
094, 095
Data error in the instru-
ment.
Reset the sensor to factory default
settings. If this fault occurs again:
Have the instrument checked by
DrägerService.
096
SIL password does not
match with PIR 7x00.
Enter password again.
Fault number Cause
Remedy
Warning
number
Cause
Remedy
101
Data logger in Stack mode
is 100 % full and is not log-
ging any more data.
Have the instrument checked by
DrägerService.
102
Data logger in Stack mode
is 90 % full.
Download the data as soon as possi-
ble and clear the data logger.
103, 106
Data error in the instru-
ment. Some dongle func-
tions like data logger,
sensor test, etc. may no
longer be available.
Reset the instrument to factory default
settings. If this error occurs again:
Have the instrument checked by
DrägerService.
104
Date or time setting not
valid.
Set date and time
105
Instrument fault.
Have the instrument checked by
DrägerService.
107
Battery for data memory
empty.
Have the instrument checked by
DrägerService.
110, 111, 112 SW dongle has been
removed without being
deactivated.
Deactivate SW dongle.
113
Alarms inhibited.
Enable alarms
164, 165
Sensor warm-up phase
has not ended. Increased
measurement error must
be expected.
Wait until the sensor has warmed up.
Do not calibrate before sensor is fully
warmed-up.
167, 170
Calibration interval
expired.
Recalibrate the instrument.
171
1)
1) Polytron
®
8100 EC only
Negative reading. Value
below range minimum.
Recalibrate zero point.
163
Low EC sensor vitality
Replace sensor
Warning
number
Cause
Remedy
163
Sensor temperature is very
high.