
Sterilization with ethylene oxide
Only the sensor cell of the O
2
sensor can be sterilized in
this way –
no other parts!
Maximum temperature: 50
°
C.
Flash-off time
: at least 24 hours at 50
°
C with 60-fold
exchange of air. This impairs the service life of the
O
2
cell!
Steam sterilization at 121 or 134
°
C
Cato parts which can be sterilized in this way are listed
in the table on page 87.
Steam sterilization must be preceded by sterilization with
moist heat and/or wipe cleaning. All parts must be dry!
Steam sterilization is only required after treating an
infectious patient!
Minimum exposure times: 20 minutes at 121
°
C and
10 minutes at 134
°
C. Parts should preferably be
sterilized at 134
°
C since less time is required for the
procedure.
Care
Care methods
Sterilization with ethylene oxide
Steam sterilization at 121 or 134
°
C
86