
Then:
●
Put inspiratory O
2
sensor back and calibrate
or
●
Put O
2
sensor for sidestream O
2
measurement back
and calibrate.
●
Put O
2
sensor back,
and
●
screw sampling hose back onto water trap.
Manual ventilation function
Machine in Manual/Spontaneous mode:
1
Press »
MAN/SPONT
«.
2
Pressure limiting valve (APL) set to MAN.
●
Connect lung simulator, exercise thorax or breathing
bag to Y-piece.
3
Set fresh gas flow.
2
Set maximum ventilation pressure between 5 and
70 mbar on APL valve – turn valve head for this
purpose.
●
Compress breathing bag.
●
Compare pressure indication on monitor with setting
on pressure limiting valve.
If excess pressure has to be relieved quickly:
2
Press vertically down on the lever of the pressure
limiting valve.
Automatic ventilation function
●
Connect lung simulator, exercise thorax or breathing
bag to Y-piece.
4
Set fresh gas.
5
Press »
IPPV
«.
6
Press rotary control.
7
Piston movement is indicated on the bar graph.
The machine starts with the ventilation parameters set
upon delivery (or programmed to customer's
requirements by DrägerService) (see page 39).
●
Lung simulator inflates regularly.
●
Pressure profile is displayed on the monitor.
●
Volume measurement yields plausible values.
105
Check operational readiness
Manual ventilation function
Automatic ventilation function
1
2
3
7
5
6
4
03437224
03537224