Messages, cause and remedy
Messages on the ventilator
Messages on the ventilator
Messages
Cause
Remedy
????
Internal electronic fault.
Call DrägerService.
compliance test
The hose compliance is being tested during the
leakage test in »standby« mode.
Equipment fault
Attempt to cancel self-test when
–
a malfunction has been signalled beforehand or
–
self-test has already been broken off 10 times!
(Mechanical ventilation is only possible when self-
test has been completed successfully!)
This message precedes the message »Malfunction
No. XXX«.
Acknowledge to cancel alarm
tone. Restart.
For information.
Note malfunction number and
call DrägerService.
GB. Version ....
Message displayed at the beginning of the self-test. Software version number is
indicated when switching on.
Inform DrägerService of version
number when reporting
malfunctions.
Leak test IPPV
The IPPV leakage rate is being determined during
the leakage test in »standby« mode.
Paed. hoses !
When changing the V
T
setting from values above
200 mL to less than 200 mL.
Use infant hoses. Run leakage
test to measure new compliance,
otherwise limited accuracy!
Press rotary control to confirm!
piston test
The system occasionally runs a self-test to detect
major leaks and faults in the inspiration valve.
Pressure release
A static excess pressure continued for too long
after AutoWakeUp. The system is relieved.
Switch ventilator to »
standby
«.
Disconnect fresh gas.
Restart
Transient system fault.
Power supply briefly interrupted.
Manual ventilation is not possible
for approx. 10 seconds while
the message »Warmstart« is
displayed; wait for 10 seconds,
the ventilator then continues with
the former settings. Check
parameter settings. Set up a
substitute system if warmstart is
repeated. Call DrägerService.
Self-test
Self-test has been started.
Self-test ... x discont
Message displayed after cancelling the self-test.
Only possible 10 times!
A complete self-test
must
then be performed.
Operating mode can be
selected. Skipping calibration
functions and internal alignment
procedures can impair the
accuracy of measurements!
Standby
Machine on standby.
Supply pressure?
Pressure supply inadequate.
Correct pressure supply.
Use spare cylinders if necessary.