Configuring monitor functions in
»standby«
Invoking the configuration menu
●
The configuration menu is called up by pressing the
softkey »
config
«.
Configurations which are set under »
defaults
« in
»
standby
« mode remain active whenever the machine
is switched on again.
All other configurations, on the other hand, only apply
temporarily.
The screen shown alongside appears on the monitor:
The
–
defaults
,
–
calibration
,
–
alarms
and
–
anaesthetic gas
can be altered with this menu.
These settings are always called up and changed in the
same way:
●
Turn rotary control
to displace the cursor frame.
●
Press rotary control
to confirm the selection made with the cursor frame.
●
Fields with grey background
represent the currently valid settings.
●
Fields with white background
show the menu steps via which the momentary field
was reached.
●
The arrow
(
z
)
means return to the preceding menu level.
Changing default values
●
Turn rotary control on monitor until the cursor frame is
located over »
defaults
«.
●
Press rotary control.
The operator is then prompted to enter a four-digit
password (specified during the initial training) in order
to prevent unauthorized changes to the basic functions.
This function can be deactivated by DrägerService.
●
Use the rotary control to select and confirm individual
numbers from the line displayed below. The password
is represented by asterisks (
*
*
*
) below the line.
When the correct password has been entered, the
cursor frame jumps to the selection field for the
default values.
51
Monitor functions
Configuring monitor functions in »standby«
Invoking the configuration menu
Changing default values
erase
trend
Select ventilation mode at the ventilator
or switch on screen by pressing
Standby
Cato
SW-Version 2.05
02-05-00
config.
Standby / Configuration
✓
anaesth. gas
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
no
anaesth. gas
warning
default
✓
calibrating
O
2
-sensor
21 Vol.%
flow sensor
Ventilator
start up test
more
✓
Alarms inactive!
defaults
pulse to.
0 1 2 3 4
5
6 7 8 9
alarm tone
1 2 3
4
5 6 7 8 9
mode
adult
Neo.
parameters
record
interfaces
alarm limits
curves
basic configuration
Standby / Configuration
Alarms inactive!
✓
anaesth. gas
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
no
anaesth. gas
warning
default
✓
calibrating
O
2
-sensor
21 Vol.%
flow-sensor
Ventilator
start up test
more
✓
defaults
These default settings are
password protected. The default
settings are active on start-up.
To change the default settings
enter the password:
0 1 2 3 4 5 6 7 8 9
* * *
00737225
02628971
02728971