Anaesthetic gas scavenging system (AGS)
●
Connect the transfer hose to the scavenging adapter
– first time only, it then remains in place.
1
Insert the scavenging adapter in the breathing system
from below until it engages.
2
Route the transfer hose round the equipment to the
rear.
3
Plug the transfer hose to the port on the collecting
system.
4
Connect the suction hose (sampling hose) to the port
on the collecting system.
5
Connect the anaesthetic scavenging connector to the
sampling hose.
6
Make sure that the second connector on the collecting
system is sealed with the screw plug.
●
Follow the specific Instructions for Use of the
anaesthetic gas scavenging system.
Anaesthetic agent vaporiser
The illustration shows a Vapor 2000
●
If using the double connector for two Vapors, the
automatic Vapor identification system is disabled.
●
Only use Vapors listed in the Order List.
●
Follow the specific Instructions for Use of the Vapor.
●
For each anaesthetic agent, only use the specified
Vapor.
7
Always insert the sealing plug and
8
secure with lever.
9
Engage control dial at
»0«
when no fresh gas is set.
10
Always secure the Vapor with the locking lever (turn
lever to the left as far as it will go).
19
Preparing for use
Connecting the equipment
Anaesthetic gas scavenging system (AGS)
Anaesthetic agent vaporiser
10
9
8
7
0
5
6
3
4