Breathing system
●
Remove flow sensor.
●
Unscrew pressure limitation valve APL.
●
Dismantle breathing valves –
valve discs must be handled with care.
●
Replace measured gas return hose if contamination
is possible.
1
Release the five quick-release screws by turning
through a quarter-turn with a 6 mm hexagon key.
Remove upper part.
2
The two valves are only unscrewed if soiled.
●
Empty soda lime container (see page 45).
Dispose of with household refuse (but not when
treating infectious patients!).
Piston pump
3
Release the two quick-release screws by turning
through a quarter-turn with a 6 mm hexagon key.
4
Remove cylinder head.
5
Completely unscrew the central screw on the piston
(no quick-release).
6
Remove piston and roller diaphragm.
●
The remainder of the piston pump does not come in
contact with breathing gas and does not require
sterilization!