Care methods
Care
Care methods
Care schematic
84
Care schematic
Machine without parts in contact
with patient gas
Manual preliminary
cleaning
In automatic disinfection
unit ANDA / Purfactor
Manual, scrubbing,
wipe disinfection
Care, rinsing and drying
Steam sterilization at 134
°
C / 121
°
C
Reassemble machine
Assembly
Manual, scrubbing, wipe and
instrument disinfection
Dismantle parts in contact
with patient gas
Strip down machine
Use
Check machine according to
checklist immediately before use
Care intervals
A number of steps can be omitted if microbial filters are used in the inspiratory and expiratory lines.
The following table is intended as a guide; it does not replace the directions of the hospital's hygiene officer!
Non-infectious patients
Infectious patients
Component:
with microbial filters:
without microbial filters:
with or without
microbial filters:
Tube, mask
after every patient
after every patient
after every patient
Y-piece
after every patient
after every patient
after every patient
Both breathing hoses
after every patient
after every patient
after every patient
Temperature sensor with lead
after every patient
after every patient
after every patient
T-piece and filter of measured gas sampling system
after every patient
after every patient
after every patient
Flow sensor downstream of expiration nozzle
daily
after every patient
after every patient
Breathing bag with hose
daily
after every patient
after every patient
Breathing system
weekly
after every patient
after every patient
Soda lime container
daily
after every patient
after every patient
Anaesthetic gas scavenging hose
weekly
weekly
after every patient
Piston pump
weekly
after every patient
after every patient
Machine ready for use
Check operational readiness
in acc. with
Instructions for Use
In automatic dis-
infection unit Aseptor