This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections
and maintenance work carried out by the manufacturer.
6132.300 Babytherm 8004/8010
Test List
Page 7
K6132300TL1.fm
24.03.06
Copyright
reserved. Rep
roduct
ion
allow
e
d f
o
r non-commer
c
ia
l purposes only.
4.3
Checking the Patient Leakage Current
•
Plug the power cord of the Babytherm into the VDE tester (see
Figure below).
•
Plug the power cord of the VDE tester into the mains power outlet.
•
Switch on the VDE tester.
•
Switch on the Babytherm.
Abb. 2:
Patient Leakage Current Test Setup
Legend
The patient leakage current should be equal to or lower than 4.0 mA.
1
Short-circuit in mains connector (connector item no.
7900765)
2
Skin temperature sensor short circuit (item no. 7910484)
3
Unit earth (earthing stud)
4
Socket SL (previously COM) on portable tester
5
Socket ABL on portable tester