Refer to figure 3.1 for dimensional requirements for wall mount. Be sure the CL4/CL8 is properly secured to
the wall.
3.3 Sample Tubing
Sample tubing may be connected to the CL4/CL8 on the top or left side of the CL4/CL8 (see section 3.3.1). All
sample tubes are 1/4” OD x 3/16” ID Teflon FEP (300 ft max length) which may be purchased from DOD
Technologies, INC (See Appendix A).
Fully depress each sample tube into the proper hole when attaching. To detach the tube, push on the collet
and pull the tubing out
3.3.1 Sample Manifold Relocation
The quick connect system used to attach sample tubing and exhaust may be located on the top or left side of
the machine. To move the manifold, Follow the procedures in section 6.2 to open the service door.
•
Remove the four (4) mounting screws shown in figure 3.5 to detach the manifold.
•
Remove the 4 lock nuts holding the cover plate over the unused manifold opening
•
Remove the cover plate
•
Align the manifold inside the CL4/CL8 with the holes in the new position
•
Replace the screws holding the manifold in place and tighten accordingly.
•
Replace the cover plate over the unused opening and secure with the four lock nuts.
Figure 3.5
Sample tubing may be purchased from DOD technologies (see appendix A)
3.3.2 End of line particulate filters
End of line particulate filters must be installed on all 4/8 sample lines at all times to prevent damage to the
unit. Unused lines must either be plugged or have a filter installed. Filters require regular maintenance
–
see
chapter 6.
IMPORTANT:
All sample tubing used with the CL4/
CL8 must be 1/4”OD x 3/16”
ID FEP Teflon. Use of any other tubing may damage the CL4/CL8 and/or cause
inaccurate gas concentration readings.