64
DC-ITD-CL4andCL8MAN01.A
www.dodtec.com
NOV 2018
815-788-5200
CL4 & CL8 Manual
Figure 4
Figure 5
9)
Enter 1225 for the password then press
“
Enter
”
.
10)
You should see the calibrate optics screen (Figure 5), then make sure the gate is closed and press
“
Calibrate
”
.
11)
If the DAC values are 140 or below you are done. Go to
“
Main Menu
”
and clear the original faults and
re-enter analysis.
12)
If the DAV values are not lower than 140, follow the advanced cleaning instructions below.
H.2 Advanced Optics Cleaning
1)
Repeat above steps 1-5.
2)
Using a cotton swab and alcohol (you may have to trim the cotton swab to fit) clean each channel that
has a DAC above 140. (Figure 6)
Figure 6
3)
Repeat above steps 7-11
4)
If the DAC values still have not improved contact CL4/CL8 technical support at (815)-788-5200