55
DC-ITD-CL4andCL8MAN01.A
www.dodtec.com
NOV 2018
815-788-5200
CL4 & CL8 Manual
Appendix D
–
System Event Messages
COLOR CODING
RED
Gas Alarm
ORANGE
Service Fault (critical)
YELLOW
Maintenance (non-critical)
BLUE
Information Message
GREEN
Status (during analysis)
Event Code
Message
Corrective Action
30000
System Communication Failure
Service required
30001
All Points Are Disabled
Enable 1 or more points - sections 4.2.a
30002
Low flow multiple points
Check pump, gate, or adjust flow
30003
Gate Close Fault
Gate home switch stuck or gate not moving
30004
Gate open Fault
Check gate home switch and/or mechanism
30005
High background Fault
Check/replace tape
30006
I/O Failure - Check wiring
Check RS-485 wiring
30007
No Gas Family Selected
Service required
30008
Pump failure
Check pump wiring/flow
30009
Tape Advance Fault
Tape switch not activated before timeout
30010
Optic 1 Communication Error
Check Wiring or Optic 1
30011
Optic 2 Communication Error
Check Wiring or Optic 2
30012
High Background Tape Advance Fault
Check/replace tape
30013
LED Failure Optic 1
Possible Tape alignment problem
30014
Warning : Default values restored
Reconfigure & Calibrate
30015
Pump Timeout Error
Pump failure or system problem
30016
Tape Advance Problem
Service required
30017
Optic 1 LED Failure
Service required
30018
Optic 2 LED Failure
Service required
30019
Optic Supply Voltage Failure
Service required
30020
Optic reference validation fault
Service required
30021
LED Failure Optic 2
Service required
30022
Complete Optic Low Flow all Pt 1-4
Check gate closure
30023
Complete Optic Low Flow all Pt 5-8
Check gate closure
30024
LED Calibration error
50000
Chemlogic tape supply low
Change Tape
50001
Pt X High Flow
Flow Adjustment - Section 4.2.d
50009
Pt X Low Flow
Flow Adjustment - Section 4.2.d
60000
8 Hour - conc. detected
60010
Last analysis - Conc. detected
60011
Alarms & Faults Reset
60012
Error Transferring TWA/Analysis Summary
60013
Flow Fluctuation
60014
Maximum # of Ref fluctuations