Technical Specifi cations
Supply voltage
230V AC
Nominal voltage
max. 4 A
Frequency
50/60 Hz
Fuse
T 6.3 A H 250 V primary/ time-delay
Maximum power consumption
800 VA
Device class according to the 93/42/EEC Directive
IIa
Protection class
Device of protection class I
Application parts
Type B application parts
Contamination level
2
Over voltage category
II
Power cable
3x1,5 mm2
Suction control lines to the suction device
5x1,5 mm2
Potential equalisation
1x 4 mm2
Relay control line optional special function
3x1,5 mm2
Free end electrical cables above fl oor
500 mm
Fuse for domestic installation
Circuit breaker: 16 A medium-lag
Recommendation: circuit breaker type C
Degree of protection against ingress of water
Ordinary device (without protection against water ingress)
The foot control is drip-water protected according
to protection class IPX8
The motors of the treatment unit are designed for intermittent operation in accordance with the dental
treatment method.
Driving motors for patient chair and backrest: duty cycle (max. 25 s „ON“ / 400 s „OFF“)
Permanently connected device. In order to avoid the risk of electric shock, this device may only be
connected to a power supply with an earth conductor.
Transport and storage conditions
Ambient temperature
30 to +70 °C
Relative humidity
10 to 80 %
Atmospheric pressure
500 hPA to 1060 hPa
Operating conditions
Quality and load-bearing capacity of the fl oor
The fl oor must be level and horizontal according to EN
18202. Unevenness of the fl oor along the total length of
the chair base up to 2 mm is acceptable. The minimum
load-bearing capacity of the fl oor must be 0.5 N/cm2
(equivalent to approx. 500 kg/m2).
Ambient temperature
10 to 35 °C
Relative humidity
15 to 80 %
Atmospheric pressure
700 hPA to 1060 hPa
Installation site
≤ 3,000 m above sea level
The treatment unit is not suitable for operation in
hazardous areas.
Weight
L2-Eco
max. 230 kg
Содержание L2-ECO
Страница 32: ...32 Cleaning and Disinfection of the Suction Handpieces...
Страница 33: ...Cleaning and Disinfection of the Suction Handpieces 33...
Страница 53: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Страница 66: ...3 Product description Motor EM 12 L Motor shealt LED Connection for instruments as per ISO 3964 O rings Seal 14...
Страница 67: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Страница 97: ......
Страница 99: ...Instructions for Use Foot control S NW S N2 S N1...
Страница 130: ...32 Manufacturer s declaration...
Страница 131: ...33 Manufacturer s declaration...
Страница 132: ...34 Manufacturer s declaration...
Страница 133: ...35 Manufacturer s declaration...
Страница 135: ...1 Minilight Installation and use manual...
Страница 146: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Страница 161: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Страница 162: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 189: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Страница 190: ......
Страница 207: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Страница 223: ......
Страница 225: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Страница 226: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 253: ......
Страница 266: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...