26
7117100018L30 1712V002
Troubleshooting
Fault
Probable cause
Solution
Yellow display is on
GREEN LED illuminates
Audible signal melody
sounds
Amalgam collecting container is
95% full
i
Change the amalgam collecting
container�
Float sensor dirty or blocked
i
If this display occurs repeatedly
even when the collecting con-
tainer is empty, check that the
float sensor can move freely�
Yellow display is on
Red display flashes
Audible signal melody
sounds
Amalgam collecting container is
100% full
i
Change the amalgam collecting
container� Audible signal can no
longer be switched off�
Float sensor dirty or blocked
i
If this display occurs repeatedly
even when the collecting con-
tainer is empty, check that the
float sensor can move freely�
Waster water line/siphon trap
dirty
i
Clean the waste water line/si-
phon trap� *
The RED and GREEN dis-
plays flash alternately
Audible signal
Motor is dirty or defective
i
Check motor for smooth run-
ning; replace the centrifuge if
necessary� *
i
Replace the device� *
Contact problems at X9
i
Plug in the connector correctly� *
i
Replace the PCB main board
and connector on the motor� *
Orange LED flashes
Audible signal
Press the service key briefly to
switch off the audible signal
Amalgam collecting container not
correctly in position
i
Switch OFF the device�
i
Insert the amalgam collecting
container in the correct position�
i
Switch ON the device�
Float sensor missing
i
Insert the float sensor� *
Water accumulating in the
spittoon
Coarse sieve in the fluid inlet
blocked
i
Clean the coarse sieve�
Outlet ineffective or not vented
i
Check or retrofit the ventilation� *
Suction power too weak or
interrupted
Coarse sieve is blocked on the
inlet of the aspiration
i
Clean the coarse sieve�
Place selection valve not or
incompletely open
i
Check the control voltage� *
i
Clean the place selection valve�
*
Device running continu-
ously
Float sensor blocked in water
start position
i
Clean the float� *
i
Free up the float sensor linkage
so that it can move freely� *
Start signal at the signal input (X2)
i
Check the control voltage� *
Waster water line/siphon trap
dirty
i
Clean the waste water line/si-
phon trap� *
Noise at the spittoon
Outlet ineffective or not vented
i
Check or retrofit the ventilation� *
EN
Содержание L2-ECO
Страница 32: ...32 Cleaning and Disinfection of the Suction Handpieces...
Страница 33: ...Cleaning and Disinfection of the Suction Handpieces 33...
Страница 53: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Страница 66: ...3 Product description Motor EM 12 L Motor shealt LED Connection for instruments as per ISO 3964 O rings Seal 14...
Страница 67: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Страница 97: ......
Страница 99: ...Instructions for Use Foot control S NW S N2 S N1...
Страница 130: ...32 Manufacturer s declaration...
Страница 131: ...33 Manufacturer s declaration...
Страница 132: ...34 Manufacturer s declaration...
Страница 133: ...35 Manufacturer s declaration...
Страница 135: ...1 Minilight Installation and use manual...
Страница 146: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Страница 161: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Страница 162: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 189: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Страница 190: ......
Страница 207: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Страница 223: ......
Страница 225: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Страница 226: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 253: ......
Страница 266: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...