7117100018L30 1712V002
11
Product description
As soon as no further fluid is fed to the amal-
gam separator, e�g� when the suction hose is
placed back in the hose manifold, the centrifuge
drum is switched off after a short delay time�
This switch-off brakes the motor, as a result
of which the ring of water, which continues to
rotate due to inertia, rinses the separated parti-
cles out of the centrifuge (8) downwards into the
collector vessel�
The separated amalgam particles form a sed-
iment in the replaceable collector vessel� The
level of fluid in the collector is regulated by the
pump cone so that the risk of fluid escaping
when the collector vessel is changed can be
avoided�
5.6 Sediment level measurement
The fill level in the collector vessel (12) is
checked by a float sensor (13) every time the
main power switch is switched on�
The centrifuge motor starts, fluid is transported
via the pump cone to the centrifuge drum (8)
and provides a constant level of fluid (under-
side of the cone pump) in the collector vessel�
The float sensor sinks� Two light barriers (9a)
and (9b) measure the fluid level� Once the level
reaches 95% in the collector vessel, this is dis-
played on the display panel�
5.7 Operating problems
If the unit is not ready for operation due to a
fault, this will be indicated on the display panel
via illuminated LEDs and an audible signal�
5.8 Service key
On the display panel there is a service key that
can be used to switch off the audible signal in
the event of a fill level warning or if a fault mes-
sage is indicated� This button can also be used
to start the device manually� To do this, press
the button for longer than 2 seconds until the
drive motor starts up�
EN
Содержание L2-ECO
Страница 32: ...32 Cleaning and Disinfection of the Suction Handpieces...
Страница 33: ...Cleaning and Disinfection of the Suction Handpieces 33...
Страница 53: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Страница 66: ...3 Product description Motor EM 12 L Motor shealt LED Connection for instruments as per ISO 3964 O rings Seal 14...
Страница 67: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Страница 97: ......
Страница 99: ...Instructions for Use Foot control S NW S N2 S N1...
Страница 130: ...32 Manufacturer s declaration...
Страница 131: ...33 Manufacturer s declaration...
Страница 132: ...34 Manufacturer s declaration...
Страница 133: ...35 Manufacturer s declaration...
Страница 135: ...1 Minilight Installation and use manual...
Страница 146: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Страница 161: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Страница 162: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 189: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Страница 190: ......
Страница 207: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Страница 223: ......
Страница 225: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Страница 226: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 253: ......
Страница 266: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...