- 106 -
1
1
.
.
1
1
0
0
O
O
p
p
t
t
i
i
o
o
n
n
s
s
1.
10.13 Pressure Measuring Instrument
•
Baratron
•
Manufacturer: MKS (
www.mksinstruments.de
)
•
Designated use: mainly for processes with CF
4
1.10.14 Power Display
•
Shows the actual power of the generator.
•
Analog display
1.10.15 Timer LT4H – Instead Of Standard Timer
•
To set process time
•
Manufacturer: NAIS, Type LT4H
Operation Of Timer:
1. The timer turns on automatic at running machine, when the pressure setpoint is
reached (0.4 mbar).
2. The timer setting is installed into minutes by standard (max. 999.9 minutes), if you
want to change the settings into hours or seconds, you have to open the machine
(left or right side wall). There you find the back of the timer. Then follow the
instructions in the manual of the timer.
3. To adjust the time, press the arrow keys up or down. The numbers in front of the
comma are ‚real’ numbers, the place behind the comma is indicated in tenth. (e.g.:
15.5 on the timer display means: 15 minutes and 30 seconds).
4. To confirm the process time press ‚Reset’.
5. At starting the generator, the time will run down. That means you see how many
time to the end of the process is left.
6. The generator stops / turns off automatic after course of time.