PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 64
3.4.6.1
<1LIMITS> (CUSTOM CONTROL PARAMETERS)
PURPOSE
To view active pressure control parameters and/or set custom control
parameters.
PRINCIPLE
See Section 3.4.6,
PRINCIPLE.
OPERATION
[SETUP], <6control>, <1limits> views and/or sets limits for the active control
mode only. To set limits for a control mode other than the active control mode,
select that control mode first using [SETUP], <6control>, <2mode>.
To view active control parameters and/or set custom control parameters, press
[SETUP]
,
<6control>
,
<1limits>
from the MAIN RUN screen. The display is
depends upon the active control mode.
To reset default control parameters for a range and control mode, select the
control mode using [SETUP], 6control, 2mode.
Custom control parameters are range specific. Changes made in one range are
stored for that range and are recalled when that range is made active. Changes
made in one range do not affect any other range.
When custom control parameters are active, a <C> is appended to the control
mode indicating character (<D> for standard dynamic, <d> for hi volume dynamic,
<S> for static, <M> for monotonic, <R> for rate) in the middle of the bottom line of
the MAIN RUN screen.
Dynamic (standard and hi volume), Static and PG Control Modes
When dynamic or static control mode is active, the
[SETUP]
,
<6control>
,
<1limits>
display is:
1.
Active Q-RPT designator.
2. View/edit field of hold limit value in active
pressure unit of measure. Shows active hold
limit.
3. Active control mode (
<D>
for standard
dynamic,
<d>
for hi volume dynamic,
<S>
for
static).
Hold limit: IH
0.010 MPa D
Содержание PPCH-G
Страница 10: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 96: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 86 N NO OT TE ES S...
Страница 154: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 144 N NO OT TE ES S...
Страница 158: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S...