PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 98
DP
Purpose
Decrease the pressure slowly by the requested amount.
Command “DP
n
”
Classic “DP=
n
”
Arguments
n:
The decrease in pressure desired (current pressure units). This can be from
0 to 2 % FS of the active range.
Remarks
The down control valve will be used to decrease the pressure the requested amount at a rate
of 0.2 % FS/sec. The PPCH-G controls pressure up to the target and then stops, so the
change in pressure will be approximate.
Example
(enhanced)
Cmd sent:
Reply:
“DP 20”
“20 kPa”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“DP? 20”
“20 kPa”
Example
(classic)
Sent:
Reply:
“DP=20”
“20 kPa”
Errors ERR#
6:
The
n
argument is not within given limits.
See Also
“IP”, “DS”, “IS”, “DF”, “IF”
3.4.3, 3.1.3
DRV
n
Purpose
Read or set the status of external electrical drivers.
Command
Query
“DRV
n
x
”
“DRV
n
?”
Classic “DRV
n=x
”
“DRV
n
”
Arguments
n:
x:
The driver to operate. This can be from 1 to 8.
The state to change the driver to; ‘0’ to de-activate it, ‘1’ to activate it.
Remarks
The PPCH-G control has eight optional external drivers.
Example
(enhanced)
Cmd sent:
Reply:
“DRV1 1”
“1”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Reply:
“DRV2? 1”
“1”
Example
(classic)
Sent:
Reply:
“DRV1=1”
“DRV1=1”
Errors ERR#
6:
The
n
or
x
arguments are not within given limits.
See Also
3.4.7
DS
Purpose
Decrease the pressure slowly.
Command “DS
n
”
Classic “DS=
n
”
Arguments ‘0’
‘1’
Stops the pressure generation.
Starts a descending generation at a slow fixed rate..
Remarks
The down control valve is used to establish a fixed rate of 0.2 % FS/sec. The valve is closed
as the pressure approaches the target.
Example
(enhanced)
Cmd sent:
Reply:
“DS 1”
“1”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“DS? 1”
“1”
Example
(classic)
Sent:
Reply:
“DS=1”
“DS=1”
Errors ERR#
6:
The
n
argument is a ‘0’ or a ‘1’.
See Also
“IF” , “DF”, “IS”, “IP”, “DP”,
3.2.1
ERR
Purpose
Read the new available error message from the Error Queue.
Query “ERR?”
Classic “ERR”
Remarks
This program message obtains additional details about an error that has occurred. If the user
receives an “ERR# nn” reply, or the enhanced mode is enabled using the IEEE-488 interface
and an error has been detected, the error is put into a “FIFO” (first in first out) error queue.
The “ERR” program message pulls and replies the oldest error message available. “OK” is
replied if there are no error messages left.
Example
(enhanced):
Query sent:
Reply:
“ERR?”
“Numeric argument missing or out of range”
Example
(classic)
Sent:
Reply:
“ERR”
“Numeric argument missing or out of range”
See Also
4.4.2
Содержание PPCH-G
Страница 10: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 96: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 86 N NO OT TE ES S...
Страница 154: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 144 N NO OT TE ES S...
Страница 158: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S...