PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 44
3.3.6.2
QDUT AUTOTEST
OPERATION
The QDUT AutoTest AutoRanges the PPCH-G (see Section 3.3.4) based on the
span and tolerance of a DUT before running a test sequence. QDUT AutoTest
also calculates DUT in or out of tolerance conditions and collects data while
running the test. See Section 3.4.5 for information on QDUT AutoTest data files.
To run a QDUT Test, press
[AutoTest]
and select
<2QDUT>
.
The display is:
Run current QDUT
AutoTest: 1yes 2no
This selection provides the choice between running the last QDUT Test,
including AutoRanging the PPCH-G based on the previously entered DUT
characteristics or setting up a new QDUT Test definition. Select
<1yes>
to rerun
the last QDUT Test (see below in this Section). Select
<2no>
to define a new
QDUT Test.
When
<2no>
is selected from the Run Current QDUT Test menu, the set-up/edit
routine is accessed. The information needed to AutoRange PPCH-G based on
the DUT characteristics is then collected (see Section 3.3.4).
The first display is the selection of the
pressure control mode in which the test is
to be run (see Section 3.4.6).
Control: 1dynamic
2static 3monotonic
Select the pressure control mode that is to be used by the PPCH-G when running the
test points..
The next display is the selection of the
measurement mode of the DUT (see
Section 3.3.3). If only gauge mode is
available, this display is skipped.
DUT measurement mode
1abs 2gage
Select the measurement mode of the DUT that is being tested. If the desired
measurement mode is not available, the QRPTs available to PPCH-G do not support
the measurement mode or it is turned off.
The next display is to specify the unit of
measure. Select the unit of measure in
which the test is to be run. The choice of
units available can be customized (see
Section 3.5.6).
1kPa 2Pa 3MPa
4hPa 5bar 6mbar
The next display is to specify the full scale (maximum pressure) of the DUT or test.
1.
Entry field for maximum pressure of the QDUT
Test sequence.
DUT full scale:
12000 psi g
This entry determines the maximum pressure of the QDUT Test sequence. Enter the
maximum pressure and press
[ENT]
.
Sequence full scale must be achievable by a QRPT available to PPCH-G or an
error will occur when PPCH-G attempts to AutoRange to the specified full scale. (see
Section 3.3.4).
Содержание PPCH-G
Страница 10: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 96: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 86 N NO OT TE ES S...
Страница 154: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 144 N NO OT TE ES S...
Страница 158: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S...