PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 48
OPERATION
To access the HEAD function, press the
[HEAD]
function key. The display is:
1.
Entry field for head height.
2.
Test fluid currently specified for the head correction.
Edit head height:
+ 25 cm N2
Edit the head height to the desired value (positive if the DUT is above the reference level,
negative if below). Press
[ENT]
to return to the MAIN RUN screen with the new head
correction active. Press
[ESC]
to return with no changes.
The reference height of the
PPCH-G pressure measurement is
the center of the TEST port on the
PPCH-G rear `panel.
The HEAD function is NOT range specific. The HEAD ON or OFF status remains the same as
ranges are changed. Edits made to the head settings are independent of active range or Q-RPT.
When a head correction is being applied, it is indicated by <h> in the top line of the MAIN
RUN screen (see Section 3.1.1). When the head correction is zero, the <h> is not shown.
To change units of head height between inches and centimeters and to change the test
fluid, use [SPECIAL], <3head> (see Section 3.5.3).
When using external Q-RPTs in RPM4, it is recommended that the RPM4 head height be
adjusted to the reference level of the PPCH-G before connecting the RPM4. Then, the normal
PPCH-G reference level to DUT head corrections can be made for all the Q-RPTs, whether internal
or external to PPCH-G.
3.3.8
[PRIME] (CYCLE)
PURPOSE
To execute the PPCH-G test system CYCLE function. CYCLE is intended to repeatedly cycle
between a vented condition and a pressure, for example to exercise a DUT before calibrating it.
Содержание PPCH-G
Страница 10: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 96: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 86 N NO OT TE ES S...
Страница 154: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 144 N NO OT TE ES S...
Страница 158: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S...