PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 120
This register can be read using the
“*RSR?”
query, Each of these status bits can set the
RSB bit of the Status Byte Register, causing a SRQ to occur IF the RSB bit is enabled to do so.
The Standard Event Status Enable Register (
“*RSE”
program message ) determines which
of these flags are able to assert the RSB bit. The description of these bits are given as:
MEAS
Measurement ready (Bit 2)
Indicates that the PPCH-G has completed a Q-RPT measurement.
NRDY
Generation Not Ready (Bit 1)
Indicates that the PPCH-G made a transition from
Ready
to
Not Ready
as
defined by the control settings (see Section 3.2.2).
RDY
Generation Ready (Bit 0)
Indicates that the PPCH-G has reached a target pressure and is
Ready
as
defined by the control settings (see Section 3.2.2).
4.6
IEEE STD. 488.2 COMMON AND STATUS PROGRAM
MESSAGES
The PPCH-G supports a set of commands that are common to all instruments conforming to IEEE Std.
488.2. These command make it easy to perform basic function for any device that supports these
commands. These commands also cover the status reporting commands. See Section 4.5 for details on
the status registers mentioned in these commands.
Table 22.
Program message list
∗
CLS
Clear all of the status & event structures.
∗
ESE
Read or set the Standard Event Status Enable Register.
∗
ESR
Read the Standard Event Status Register.
∗
IDN
Identify the PPCH-G version, range, and serial number.
∗
OPC
Set the operation complete bit when all operations have completed.
∗
OPT
Read the list of installed PPCH-G options.
∗
RST
Reset the PPCH-G control settings to factory settings.
∗
TST
Read the power on self test status.
∗
SRE
Read or set the Service Request Enable Register.
∗
STB
Read the Status Byte Register.
∗
RSE
Read or set the Ready Status Enable Register.
∗
RSR
Read the Ready Status Register.
Содержание PPCH-G
Страница 10: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 96: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 86 N NO OT TE ES S...
Страница 154: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 144 N NO OT TE ES S...
Страница 158: ...PPCH G OPERATION AND MAINTENANCE MANUAL 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S...