6. REMOTE OPERATION
Page 157
© 2009 DH Instruments, a Fluke Company
PCAL:IH and PCAL:IL
Purpose
Read or set the Q-RPT or utility sensor calibration information for the “IH” (internal high) and
“IL” (internal low) RPTs. “IH” is equivalent to “IuH”.
Command
Query
“PCAL:IH
adder, mult, CalDate, (GaOnly)
”
“PCAL:IL
adder, mult, CalDate, (GaOnly)
”
“PCAL:IH?”
“PCAL:IL?”
Classic
“PCAL:IH
=
adder, mult, CalDate, (GaOnly)
”
“PCAL:IL
=
adder, mult, CalDate, (GaOnly}
”
“PCAL:IH”
“PCAL:IL”
Defaults
“PCAL:IH
= 0.0, 1.0, 19800101, 0”
“PCAL:IL
= 0.0, 1.0, 19800101, 0”
Arguments
Adder:
The RPT calibration adder (PA). In Pascal.
Mult:
The RPT calibration multiplier (PM) from 0.1 to 100.
CalDate:
The date of the calibration in the format “YYYYMMDD” by default. The format
“YYMMDD” is also be accepted, and the replied format is in the previously
entered format.
GaOnly
Optional “Gauge only” flag. Indicates that the an Axxx Q-RPT is set to
operate in gauge measurement mode only. This field does not apply and is
ignored for Gxxx and BGxxx Q-RPTs or utility sensors.
‘0’ Absolute, negative gauge and gauge modes are available.
‘1’ Gauge mode only.
Remarks
The user defined pressure calibration information for the specified RPT (IH or IL pressure)
can be accessed with this program message. Using this program message overwrites the
current calibration coefficients, so caution must be used. If the “
GaMode
” flag is set and the
RPT is an Axxx Q-RPT, then the RPT will only be allowed to operate in gauge mode.
Changes made using this program message take effect immediately.
The calibration information of DHI RPM4s connected to PPC4 must be adjusted directly on
the RPM4. The information cannot be adjusted through the PPC4.
Example
(enhanced)
Cmd sent:
Query reply:
“PCAL:LO 2.1, 1.000021, 20011201, 0”
“ 2.10 Pa, 1.000021, 20011201, 0” (No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“PCAL:LO? 2.1, 1.000021, 20011201, 0”
“ 2.10 Pa, 1.000021, 20011201, 0”
Example
(classic)
Sent:
Reply:
“PCAL:LO=2.1, 1.000021, 20011201, 1”
“ 2.10 Pa, 1.000021, 20011201, 1”
Errors
ERR# 6:
One of the arguments is out of range.
See Also
7.2