5. ADVANCED USER INTERFACE
Page 103
© 2009 DH Instruments, a Fluke Company
Use UUT Tolerance:
If this field is “Yes” the UUT Tolerance is used to
determine the PPC4 resolution and hold limit settings for
the resulting AutoRange. If this field is “No” the
resolution and hold limit settings for the resulting
AutoRange are determined based on the AutoRange full
scale and the active selected Q-RPT.
UUT Tolerance:
If enabled, the tolerance of the device under test, in % of
AutoRanged span, associated with the resulting
AutoRange. Used to determine control limits and to
determine pass/fail for test restuls.
[OK]
from the AutoRanger screen enters the AutoRange Confirmation screen,
where the specified Full Scale, Unit, Mode and UUT Tolerance are listed in
addition to the selected RPT. Selecting the
RPT
drop down list shows all
available RPTs that can cover the specified AutoRange. If an RPT other than the
one selected is desired, it may be chosen from this list.
Back
to return to the
AutoRanger screen.
[Esc]
or
exits to the Main Run Screen.
Note
If an external Q-RPT that should be available is not being
selected by AutoRange, use [RPT] to search for Q-RPTs and
determine whether external Q-RPTs have been properly
connected and initialized (see Section 2.3.6).
[OK]
from the AutoRange Confirmation screen creates the range and continues,
[Back]
returns to the previous screen.
[Esc]
or
exits to the Main Run Screen.
Continuing enters the Advanced Test
editing screen where test values are
listed on the left and test sequence
editing buttons are given on the right.
Note
The pressure unit for the test points is shown to the right of the
test value list and may be different from the current, active unit
of measure (see Section 3.2.10).
To edit the numerical value of a test point:
Select
the entry in the test value list and
Enter
or
to access the list.
Select
the test point to edit.
Enter
or
again to choose the point and exit to the Test Point field.
The counter shows which test point is active (test point number / total
number of points).
Edit the value of the test point with the numeric keypad.
Enter
or
to accept the edited value.
[Auto Fill]:
This function creates a test sequence of uniform steps. From the
next screen enter the following: