PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 88
Table 15.
Main Run Screen Fields and Buttons
FIELD
DESCRIPTION
PURPOSE
CONTENTS
SECTION
[Tools]
button
Access PPC4
Tools
functions
none
AutoZero
Shortcut to run AutoZero screen
(absolute measurement mode only)
5.7.5.1
Uncertainty
Shortcut to the Uncertainty settings
screen
Measured pressure uncertainty
value
7.2.7.3
Head
Shortcut to the Head settings
screen
Current pressure head value,
fluid
5.7.6.3
[AutoRange] button
Enter
the AutoRange screen
5.7.1
Target Pressure
Pressure Information
Enter
to set target pressure.
Display pressure information
depending on PPC4 function.
Label above defines contents.
Numeric value and sign
Label:
Target
(control target
pressure) or
Deviation
(deviation from
target value)
5.5.1
Measured Pressure
Display pressure measured by
active Q-RPT
Shortcut to Resolution Screen
Numerical pressure value and
sign
5.7.6.4
Ready/Not Ready
indicator
Indicate pressure
Ready/Not Ready
<Green>
:
Ready
<Yellow>
:
Near Ready
<Red>
:
Not Ready
3.2.3
Control Mode
Control Status Indicator
Indicate current control mode and
control status
Shortcut to control settings screen
Control mode:
Static
or
Dynamic
Indicator
solid:
not controlling
Indicator
animated:
controlling
5.7.8.2
Measurement Mode
Identifies measurement mode of
displayed pressure
Shortcut to measurement mode
screen
Absolute
Gauge
Negative Gauge
5.7.6.2
Unit of Measure
Identifies pressure unit of measure
in which measured pressure and
pressure information values are
displayed
Shortcut to unit of measure screen
Pressure unit of measure
abbreviation
5.7.6.1
Pressure Rate of Change
Display rate of change of pressure
in current units per second
Numerical value and sign in
current pressure units/s
none
AutoRange Range Indicator
Display AutoRange limits
Lower and upper limits of
AutoRanged range with bar
graph of measured pressure
3.2.6
5.7.1
Active Q-RPT Span Indicator
Display active Q-RPT limits
Lower and upper limits of Q-RPT
span with bar graph of measured
pressure
3.2.6
Leak Test
Shortcut to leak test screen
5.7.3.2
Purge
Shortcut to purge screen
5.7.3.1
AutoTest
Shortcut to Advanced AutoTest
5.7.2
[Settings]
button
Enter
the PPC4
Settings
menu
none
* Noted fields have pressure values expressed in current pressure units.