4. BASIC USER INTERFACE
Page 67
© 2009 DH Instruments, a Fluke Company
4.4.5
<Setup>, <Control>
PURPOSE
To set the automated pressure control mode for the active range, customize control
parameters, activate default control parameters.
OPERATION
To access the CONTROL function, use
<Setup>
,
<Control>
. The display is:
Dynamic X1H
♦Limits ♦Mode
<Limits>
to view or edit the current control limits (see Section 4.4.5.1).
<Mode>
to change the control mode and to reset custom control parameters (see Section
4.4.5.2).
The display is:
Control Mode: X1H
♦Dynamic ♦Static
The cursor is on the active control mode. Making a control mode selection activates the
control mode, sets default control parameters and returns to the Main Run Screen. To set
custom control parameters, use
<Setup>
,
<Control>
,
<Limits>
from the Main Run Screen
(see Section 3.2.13).
4.4.5.1 <Limits> (Custom Control Parameters)
PURPOSE
To view active pressure control parameters and/or set custom control
parameters.
OPERATION
To view active control parameters and/or set custom control parameters, select
<Setup>
,
<Control>
,
<Limits>
from the Main Run Screen. The display is:
1.
Active Q-RPT designator.
2.
View/edit field of hold limit value in active
pressure unit of measure. Shows active hold
limit.
3.
Active control mode (
<D>
for dynamic,
<S>
for static).
Hold Limit: X1H
0.05 kPa D
Edit the hold limit value if desired and
to view the stability.
The next display is:
1.
Active Q-RPT designator.
2.
View/edit field of stability limit value in active
pressure unit of measure. Shows active
stability limit.
3.
Active control mode (
<D>
for dynamic,
<S>
for static).
Stability Limit: X1H
0.05 kPa/s D
Edit the stability limit value if desired and
to return to the Main Run Screen
with the edited values active.
Содержание PPC4
Страница 70: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 58 Notes...
Страница 194: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 182 Notes...
Страница 226: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 214 Notes...