6. REMOTE OPERATION
Page 165
© 2009 DH Instruments, a Fluke Company
RES
Purpose
To read or set the pressure display resolution for the active range.
Command
Query
“RES
n
”
“RES”
Classic
“RES=
n
”
“RES”
Default
“RES 0.001”
Arguments
n:
The pressure display resolution in % span of the current RPT range (0.0001 to 1 % FS).
Remarks
The pressure display resolution is defined as % span of the active range. The setting is
separate for each range, and changes as the range is changed.
Example
(enhanced)
Cmd sent:
Query reply:
“RES .01”
“0.01” (No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“RES? .01”
“0.01”
Example
(classic)
Sent:
Reply:
“RES=.01”
“0.01”
Errors
ERR# 6
The argument is invalid.
See Also
3.2.19
RESET
Purpose
Reset the user’s settings to factory defaults.
Command
“RESET”
Classic
“RESET”
Remarks
The PPC4 has user settings (units, resolution, control modes, etc.) that can be reset to factory
defaults. System calibration coefficients and communications settings are not affected. The
remote “RESET” program message corresponds to the front panel “Reset Settings.” The reset
cycle takes up to 5 seconds to complete and involves a power cycle of the PPC4. Remote
communications should not take place during this period.
Example
(enhanced)
Cmd sent:
Reply:
“RESET”
“RESET”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“RESET?”
“RESET”
Example
(classic)
Sent:
Reply:
“RESET”
“RESET”
See Also
3.2.25.1
RETURN
Purpose
Start a new pressure setting sequence equivalent to the “PS” command using the last target
pressure.
Command
“RETURN”
Classic
“RETURN”
Remarks
The “RETURN” program message starts a new pressure setting sequence using the current
settings and target pressure. Pressure control continues until a new target pressure is set, the
PPC4 is put into LOCAL mode, or an “ABORT” program message is executed.
Example
(enhanced)
Cmd sent:
Reply:
“RETURN”
“1000.000 kPaa” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“RETURN”
“1000.000 kPaa”
Errors
ERR# 6
The current target pressure is invalid.
See Also
“PS“, “TP”
3.2.12