USER Manual
96
Table 11-29 Dose area product chart (At 60kV, Exposure time: 0.7 ~ 1.35 sec)
Dose Area Product (Area of 6cm x 6cm) at 60kV
Tube
Current
Entrance
Dose Value
[uGy]
Exposure Time [sec]
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.35
1mA
at 1
cm
17.01
2
19.45 22.07 24.65 27.42 29.91 32.20 33.68
at 36
cm
612.3
2
700.4 794.4 887.5 987.1
1077
1159
1213
2mA
at 1
cm
34.39
2
39.41 44.76 50.57 56.25 61.36 64.59 67.64
at 36
cm
1238
2
1419
1611
1821
2025
2209
2325
2435
3mA
at 1
cm
50.13
2
57.29 64.93 72.14 80.24 87.53 95.33 99.72
at 36
cm
1805
2
2062
2337
2597
2889
3151
3432
3590
4mA
at 1
cm
65.28
2
74.72 84.80 95.15 105.8 115.4 122.8 128.5
at 36
cm
2350
2
2690
3053
3425
3810
4156
4420
4625
5mA
at 1
cm
80.8
2
91.81 103.5 110.5 122.9 134.1 149.3 156.6
at 36
cm
2910
2
3305
3727
3977
4424
4826
5374
5638
Unit: [uGy]
Содержание ADX6000s
Страница 1: ......
Страница 9: ...USER Manual 8 1 3 Symbols marked on iRayA6s...
Страница 31: ...USER Manual 30 Figure 7 7 Press the memory button Figure 7 8 The X ray exposure setting value stored in device...
Страница 104: ...USER Manual 103 Figure 11 10 Measurement points in 1 cm from case surface 1 2 3 4 5 6 8 7 9 10 12 11 13 14 15 16 17...
Страница 110: ...USER Manual 109...