USER Manual
31
7.2.5
X-RAY exposure
After completing all settings, confirm ready status of X-RAY generating tube before
starting X-RAY exposure. If exposure ready status is complete, the EXPOSURE status
symbol of screen is displayed "Ready" and activated to green color as shown in
Figure 7.9. Also the status window at the button of the main screen indicates that
the "Tube is ready". If exposure ready status is not complete, the EXPOSURE status
indicator of screen is displayed "wating" and deactivated to gray color. Also the
status window at the button of the main screen indicates that the "Tube is not
ready". 30 seconds after one round of exposure completion, the EXPOSURE status
symbol of screen is displayed "Ready" and activated to green again.
Figure 7.10 X-RAY generating tube exposure status symbol display
If exposure ready is complete, press Exposure button on handle part. And then X-
RAY exposure status symbol as shown in Figure 7.10.
If Exposure button is pressed, the Exposure status symbol still displayed "Ready",
and it will be standing by until X-RAY output is detected.
If exposure button is pressed and turned into X-RAY exposure standby status, panel