USER Manual
27
7.2.2
X-RAY tube curren setting
iRayA6s product can change tube current in range of 1mA ~ 5mA.
In order to adjust X-RAY tube current, press tube current (mA) setting button at the
tube current of panel.
Figure 7.4 Buttons for tube current adjustment
If tube current setting button is pressed, use the arrow buttons up or down to
select the proper setting. The status changes to tube current adjustment.
Tube current: Range is 1 ~ 5mA, and it can be adjusted by 1mA increment. Using
two buttons on left among four adjustment buttons, intensity can be adjusted by
1mA increment , and two buttons on right side by 0.1mA increment.
To set tube current, press tube current adjustment button to set corresponding
value.
When exposure dose is changed, dose value (mAs) based on exposure time is
Tube current decrease of
1mA Unit Button
Tube current
Setting Button
Tube current increase of
1mA Unit Button
Tube current decrease of
0.1mA Unit Button
Tube current increase of
0.1mA Unit Button