USER Manual
74
Table 11-8 Adult technique Chart of Femur
Adult Technique Chart (Focal –Detector Distance : 70cm)
Part: Femur
Estimated Exposure Time in Seconds [sec]
No
Thickness of Femur
(cm)
75kV
1mA
2mA
3mA
4mA
5mA
1
1.5
0.1
2
3
0.2
3
4.5
0.3
4
6
0.4
5
7.5
0.5
0.3
6
9
0.6
0.35
7
10.5
0.4
8
12
0.45
9
13.5
0.5
10
15
0.55
0.38
11
16.5
0.6
0.41
12
18
0.44
13
19.5
0.47
14
21
0.5
0.39
15
22.5
0.53
0.41
16
24
0.43
17
25.5
0.45
18
27
0.47
19
28.5
0.5
0.35
20
30
0.53
0.38