USER Manual
108
The measurement results of stray radiation at 50cm from Focal spot are as Table 11-37.
Table11-37 Stray radiation at 50[cm] from focal spot
POINT A
POINT B
POINT C
POINT D
POINT E
POINT F
0 cm
5.675
4.332
2.248
2.153
2.224
2.716
20 cm
7.221
6.153
2.985
2.985
2.500
2.568
40 cm
13.650
16.323
5.000
3.778
5.478
4.222
60 cm
14.225
16.881
8.989
5.578
5.711
8.767
80 cm
16.533
17.433
7.422
7.522
8.411
5.056
100 cm
23.333
21.310
0.624
6.604
0.708
3.867
120 cm
15.311
13.567
6.123
9.100
3.500
8.600
140 cm
10.300
9.435
12.712
6.033
8.300
9.933
160 cm
13.238
15.320
11.647
10.867
7.100
8.264
180 cm
8.635
7.322
9.312
5.867
6.033
6.822
200 cm
7.332
5.218
6.773
4.467
3.467
2.556
Unit: [mR/h]
ND: Not Detected
Figure 11.14 Graph of stray radiation at 50[cm] from focal spot