USER Manual
78
Table 11-11 Pediatric technique Chart of AP Hand
Pediatric Technique Chart (Focal –Detector Distance : 70cm)
Part: AP Hand
Estimated Exposure Time in Seconds [sec]
No
Thickness of
AP Hand (cm)
60kV
1mA
2mA
3mA
4mA
5mA
1
1.5
0.05
2
3
0.09
3
4.5
0.13
4
6
0.17
5
7.5
0.22
6
9
0.26
7
10.5
0.3
0.23
8
12
0.33
0.25
9
13.5
0.3
10
15
0.33
11
16.5
0.36
0.25
12
18
0.39
0.27
13
19.5
0.29
14
21
0.31
15
22.5
0.33
16
24
0.35
0.28
17
25.5
0.37
0.29
18
27
0.3
19
28.5
0.31
0.23
20
30
0.33
0.25