70
13.1 Updating the Software
13.1.1 Main Control and Touch Display
Updating can only be done via the web interface. Registration is done as described in
section
.
Click on settings >> and then update. In order to update the main control including the
display, enter the file name and path in the input field and click "load". Alternatively you
can select the file by clicking the "Search.." button next to the input field.
After you clicked the "load" button the inscription "the update is being transfer to the
controller" appears. Wait until the update file is transfered and the AST6 is started anew.
"Updating Main Unit" appears on the display of the sequence controller AST6. After
updating the main controller the display is updated. You will recognize this by the red
loading bar on the display of the AST6 that will slowly become green.
The update is completed when the main screen with the process display is displayed on
the display.
13.1.2 Power unit
Updating can only be done via the web interface. Registration is done as described in
section
.
Click on settings >> and then update. In order to update the Power unit, enter the file
name and path in the input field and click "load". Alternatively you can select the file by
clicking the "Search.." button next to the input field.
After you clicked the "load" button the inscription "the update is being transfer to the
controller" appears. Wait until the update file is transfered and the sequence controller is
started anew. "Updating Main Unit" appears on the display of the AST6.
The update is completed when the main screen with the process display is displayed on
the display.
13.1.3 Calibrating the Touch Display
Due to aging and use it might be required that the touch display must be calibrated anew
with the following procedure:
1. Switch off the sequence controller AST6 using the main switch.
2. Touch the touch display during switching on the sequence controller AST6 and keep
pressing it.
3. After the message "touch adjustment" appears let go of the touch display.
4. Within a second you must touch the touch display for at least one second.
5. Touch as accurately as possible the point left on top and subsequently right bottom.
6. Switch off the sequence controller AST6.