ENGLISH • 7
9608 30K FSI-SLI-10S-##
CAVITRON
R
Section 6: SPS Ultrasonic Scaler Description cont'd
6.4 Foot Control Information & Operation
The Foot Control is a two-position momentary switch, which activates both ultrasonic energy and Lavage Flow at the insert tip.
O-Ring
Provides seal for handpiece
coolant.
Insert Tip
Shape and size of tip
determines access and
adaptation. Preheated
Lavage directed to tip.
Finger Grip
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Hold the handpiece in an upright position. Activate the Foot Control to bleed any air bubbles that might be trapped inside the hand-
piece. Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Fully seat insert with a gentle push-twist
motion. DO NOT FORCE.
6.3 Cavitron
®
30K
™
Ultrasonic Inserts
The many styles of Cavitron
®
30K
™
Ultrasonic Inserts are easily interchangeable
for various procedures and applications. See enclosed booklet for specific information.
Magnetostrictive Stack
Converts energy provided by the
handpiece into mechanical oscillations
used to activate the insert tip.
Insert Marking
Manufacturer, Date (YYMM),
Frequency, Type, Tip Lot.
Foot Control
released...
Both ultrasonic activation and
irrigating flow stop.
Foot Control
depressed half
way...
The ultrasonic insert tip is activated
and irrigating water flows.
Foot Control fully
depressed...
This activates the "Boost" mode.
(1st position)
(2nd position)
Section 7: Accessories
1. Steri-Mate
®
Detachable Sterilizable Handpiece
2. AC Power Cord Set
3. Cavitron
®
Ultrasonic Inserts
4. Dual position Foot Control Assembly
5. DualSelect
™
Dispensing System
*For detailed information, contact your local DENTSPLY
®
Representative or authorized DENTSPLY
®
Distributor.