background image

ENGLISH • 8

Section 8: Techniques for Use

8.1 Patient Positioning

• For optimal access to both the upper and lower arches,

the backrest of the chair should be adjusted as for other

dental procedures. This assures patient comfort and

clinician visibility.

• Have the patient turn their head to the right or left.

Also position their chin up or down  depending upon

the quadrant and surface being treated. Evacuate irrigant

using either a saliva ejector or High Volume Evacuator

(HVE).

8.2 Performing Ultrasonic Scaling

Procedures

• NOTE: Refer to the Infection Control Information card

supplied with your System for general procedures to be

followed at the beginning of each day and between

patients.

• The edges of Cavitron

®

 Ultrasonic Inserts are intention-

ally rounded so there is little danger of tissue laceration

with proper ultrasonic scaling technique. Whenever the

insert tip is placed in the patient’s mouth, the lips, cheek

and tongue should be retracted to prevent accidental

prolonged contact with the activated tip.

• Hold the empty Handpiece in an upright position.

Activate the Foot Control until fluid exits.

• Lubricate the rubber O-ring on the insert with water

before placing it into the handpiece.  Fully seat insert

with a gentle push-twist motion.  DO NOT FORCE.

• Activate the System. Hold the handpiece over a sink or

drain.  Check spray temperature to verify fluid is reaching

the working end of the insert tip. Adjust the Lavage

Control knob to ensure adequate flow for the selected

Power setting. Greater flow settings provide

cooler irrigant.

• It may be necessary to adjust lavage with the System in

“Boost” mode (Foot Control fully depressed) so

adequate fluid will be available to cool tip tooth interface.

• In general, it is suggested that a “feather-light-touch” be

used both supra- and subgingivally. The motion of the

activated tip and acoustic effects of the irrigating fluid, in

most cases, is adequate to remove even the most

tenacious calculus.

• Periodically check the Cavitron

®

 ultrasonic insert for wear

with the DENTSPLY

®

 Insert Efficiency Indicator.

• The use of a saliva ejector or High Volume Evacuator

(HVE) is recommended during all procedures.

• Set the System’s Power Adjustment knob to the lowest

power setting for the application and the selected insert.

8.3 Patient Comfort Consideration

Reasons for sensitivity

• Incorrect tip placement. Point should be directed away

from root surfaces.

• Not keeping tip in motion on tooth. Do not allow the

insert to remain in a static position on any one area of

the tooth. Change the insert’s path of motion.

• Applying pressure. Use extremely light grasp and

pressure, especially on exposed cementum.

• If sensitivity persists, decrease power setting and/or

move from the sensitive tooth to another and then

return.

Section 9: System Care

It is recommended that you perform the following maintenance

procedures to help minimize bio-film formation in the lavage

path of your Cavitron

®

 SPS

 ultrasonic scaler which could affect

the lavage flow to the ultrasonic insert tip and scaling performance.

9.1 Daily Maintenance

Start-Up Procedures at the beginning of the day:

1. Open the manual shut-off valve on the dental office water

supply system.

2. Turn the System ON using the Power ON/OFF switch. (see

illustration on page 5) Verify the Power Indicator Light is

lighted.

3. Set the Power Adjustment knob at the top of the blue scale.
4. Set the Lavage Control knob to maximum.
5. Hold the Handpiece (without an insert installed) upright over

Содержание Cavitron SPS

Страница 1: ...ically controls the tip stroke over different working conditions The Cavitron SPS scaler produces 30 000 microscopically small strokes per second at the insert s working tip This combined with acousti...

Страница 2: ...s 5 3 Electrical Requirements 5 4 Unpacking the System 5 5 Power Cord Power Connection 5 6 Foot Control Cable Assembly Connection 5 7 Water Supply Line Connection 6 SPS Ultrasonic Scaler Description 6...

Страница 3: ...rocedures 2 2 Warnings Persons fitted with cardiac pacemakers have been cautioned that some types of electronic equipment might interfere with the operation of a pacemaker Although no instance of pace...

Страница 4: ...ther areas contact your local DENTSPLY representative 4 2 GeneralInfectionControl Recommendations As with all dental procedures use standard personal protection equipment i e wear face mask eyewear or...

Страница 5: ...into an approved outlet Section 5 Installation Instructions cont d 5 6 Foot ControlCable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in un...

Страница 6: ...o position Foot Control provides the capability to increase the front panel Power dial setting When lighted indicates Foot Control is fully depressed and System is functioning in Boost mode Lavage Flo...

Страница 7: ...st motion DO NOT FORCE 6 3 Cavitron 30K Ultrasonic Inserts The many styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications See enclosed booklet fo...

Страница 8: ...depressed so adequate fluid will be available to cool tip tooth interface In general it is suggested that a feather light touch be used both supra and subgingivally The motion of the activated tip an...

Страница 9: ...lying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufacturer To clean system generously spray disinfectant solutio...

Страница 10: ...epth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for operating theatres Protection Class 1PXO Operating Environ...

Отзывы: