Section 10:  Troubleshooting

Although service and repair of the Cavitron

®

 SPS

 Ultrasonic Scaler

should be performed by DENTSPLY

®

 personnel, the following

are some basic trouble shooting procedures that will help avoid

unnecessary service calls. Generally, check all lines and connections to

and from the System, a loose plug or connection will often create

problems. Check the settings on the System’s knobs.

10.1 Troubleshooting Guide

System will not operate:

(Power Indicator Light is not lighted.)
1. Check that the Power switch is in the ON position,

and that the detachable Power Cord is fully seated in

the receptacle on back of System.

2. Check that the System’s three-prong plug is fully

seated in an appropriate AC receptacle, and that AC

current is present.

Power Indicator Light is lighted.)

Section 9: System Care  cont'd

a sink or drain. Activate the Foot Control and flush the water

line for at least 2 minutes.

6. Place a sterilized insert into the Handpiece and set the Lavage

Control knob to your preferred operating position.

Between Patients

1. Remove all ultrasonic inserts and handpiece used, clean and

sterilize.

2. Clean and disinfect the surfaces of the cabinet, Power Cord,

Handpiece Connector and cable assembly, Foot Control and

cable assembly by applying an approved non-immersion type

disinfectant solution* carefully following the instructions

provided by the disinfectant solution manufacturer. To clean

system, generously spray disinfectant solution on a clean towel

and wipe all surfaces. Discard used towel. To disinfect system,

generously spray disinfectant on a clean towel and wipe all

surfaces.  Allow disinfectant solution to air dry.

3. Connect a freshly sterilized handpiece to its matching cable

connector. Hold the handpiece over a sink or drain and flush

the water line of the unit as above at maximum water flow

for 30 seconds.

4. Place a freshly sterilized insert into the handpiece.

Shut-Down Procedures at the end of the day:

1. Remove all ultrasonic inserts and handpiece used, clean and

sterilize.

2. Turn the System OFF.
3. Clean and disinfect the surfaces of the cabinet, Power Cord,

Handpiece Connector and cable assembly, Foot Control and

cable assembly by applying an approved non-immersion type

disinfectant solution* carefully following the instructions

provided by the disinfectant solution manufacturer. To clean

system, generously spray disinfectant solution on a clean towel

and wipe all surfaces. Discard used towel. To disinfect system,

generously spray disinfectant on a clean towel and wipe all

surfaces.  Allow disinfectant solution to air dry.

4. Close the manual shut-off valve on the dental water supply

system.

*NOTE: Water-based disinfection solutions are preferred. Some

alcohol-based disinfectant solutions may be harmful and may

discolor plastic materials.

9.2 Weekly Maintenance

End of Week Procedures (when connected to a DualSelect

Dispensing System)

Follow the end of week procedures listed in the DualSelect

manual.

9.3 Water Filter Maintenance

When the water filter becomes discolored, the filter should be

replaced to prevent reduced water flow to the Cavitron

®

 SPS

Ultrasonic Scaler. A 10-pack of replacement filters is available by

ordering Part Number 90158 from your local DENTSPLY

®

distributor.

1. Disconnect the water supply hose from the water source. If a

quick-disconnect connector is attached to the end of the hose,

relieve the water pressure by pressing the tip of the connector

in an appropriate container and drain the water.

2. Grasp the fittings on either side of the filter disk and twist

counterclockwise. Remove the filter section from either side

of the water hose.

3. Install the replacement filter onto the water hose fittings. The

filter should be positioned to match up with the correct hose

fitting.

4. Hand tighten the two hose fittings in a clockwise direction.

Reconnect the water supply hose, operate the unit to bleed

the air and test for leaks.

1. Check that the Foot Control Connector is fully seated

in the Foot Control Receptacle on the back of the

System.

System operates:

 (No lavage flow at insert tip.)
1. Assure that Lavage control is properly adjusted.
2. Check that irrigant supply control valve/s (dental

office water supply) are open.

3. If connected to DualSelect

 Dispensing System, check

that fluid level  in thse selected bottle is sufficient. Make

sure valves are open when using external water source.

10.2   Technical Support and Repairs

For technical support and assistance call 1-800-989-8826 Monday

through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other

areas, contact your local DENTSPLY

®

 representative.

Содержание Cavitron SPS

Страница 1: ...ically controls the tip stroke over different working conditions The Cavitron SPS scaler produces 30 000 microscopically small strokes per second at the insert s working tip This combined with acousti...

Страница 2: ...s 5 3 Electrical Requirements 5 4 Unpacking the System 5 5 Power Cord Power Connection 5 6 Foot Control Cable Assembly Connection 5 7 Water Supply Line Connection 6 SPS Ultrasonic Scaler Description 6...

Страница 3: ...rocedures 2 2 Warnings Persons fitted with cardiac pacemakers have been cautioned that some types of electronic equipment might interfere with the operation of a pacemaker Although no instance of pace...

Страница 4: ...ther areas contact your local DENTSPLY representative 4 2 GeneralInfectionControl Recommendations As with all dental procedures use standard personal protection equipment i e wear face mask eyewear or...

Страница 5: ...into an approved outlet Section 5 Installation Instructions cont d 5 6 Foot ControlCable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in un...

Страница 6: ...o position Foot Control provides the capability to increase the front panel Power dial setting When lighted indicates Foot Control is fully depressed and System is functioning in Boost mode Lavage Flo...

Страница 7: ...st motion DO NOT FORCE 6 3 Cavitron 30K Ultrasonic Inserts The many styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications See enclosed booklet fo...

Страница 8: ...depressed so adequate fluid will be available to cool tip tooth interface In general it is suggested that a feather light touch be used both supra and subgingivally The motion of the activated tip an...

Страница 9: ...lying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufacturer To clean system generously spray disinfectant solutio...

Страница 10: ...epth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for operating theatres Protection Class 1PXO Operating Environ...

Отзывы: