Section 5: Installation Instructions

5.1 General Information

If the installation of  your Cavitron

®

 SPS

  System is performed

by someone other than trained DENTSPLY

®

 Distributor

personnel, care should be taken to observe the following
requirements and recommendations.

5.2 Water Line Requirements

• The System’s water supply line is factory installed. Do

not disconnect from the System.

• Incoming water supply line pressure to the System must

be 25 psi (172 kPa minimum) to 60 psi (414 kPa

maximum).  If  your dental water system’s supply line

pressure is above 60 psi, install a water pressure

regulator on the water supply line to your Cavitron

®

SPS

 System.

• A manual shut-off valve on the dental water system

supply line should be used so that the water can be

completely shut-off when the office is unoccupied.

• A filter in the dental water system supply line is

recommended so that any particles in the water supply

will be trapped before reaching the System.

• After the above installations are completed on the dental

water supply system, the dental office water line should

be thoroughly flushed prior to connection to the

System.

Section 4: 

 

Infection Control

4.1 Infection Control Information

Reference Card

For your convenience, an Infection Control Information

reference card has been included with your Cavitron

®

 SPS

System. Additional cards can be obtained by calling Customer

Service at 1-800-989-8826 Monday through Friday, 8:00 AM

to 5:00 PM (Eastern Time). For other areas, contact your local

DENTSPLY

®

 representative.

4.2 General Infection Control

  Recommendations

• As with all dental procedures, use standard personal

protection equipment (i.e., wear face mask, eyewear,

or face shield, gloves and protective gown).

• For maximal operator and patient safety, carefully follow

the Infection Control Information procedures detailed

on the reference card accompanying your System.

• As with high speed handpieces, and other dental

devices, the combination of water and ultrasonic

vibration from your Cavitron

®

 SPS

  System will create

aerosols. With proper technique, much of the Cavitron

®

SPS

  System’s aerosol dispersion can be effectively

controlled and minimized. Please carefully follow the

procedural guide lines in this manual regarding the use

of your System.

4.3 Water Supply Recommendations

It is highly recommended that all dental water supply systems

conform to applicable  CDC (Centers for Disease Control and

Prevention) and ADA (American Dental Association) standards,

and that all recommendations be followed in terms of flushing,

chemical flushing, and general infection control procedures. See

sections 4.2 and 9.

5.3 Electrical Requirements

Refer to Section 11: Specifications.

5.4 Unpacking the System

Carefully unpack your Cavitron

®

 SPS

 System and verify that all

components and accessories are included:

1. Cavitron

®

 SPS

 System with water supply line and

handpiece cable connector assembly factory installed.

2. Detachable AC Power Cord set.
3. Steri-Mate

 Detachable Sterilizable Handpiece.

4. Dual position Foot Control Assembly.
5. Cavitron

®

 Ultrasonic Inserts (quantity optional).

6. Efficiency Indicator for Cavitron Inserts.
7. Literature Packet.

Содержание Cavitron SPS

Страница 1: ...ically controls the tip stroke over different working conditions The Cavitron SPS scaler produces 30 000 microscopically small strokes per second at the insert s working tip This combined with acousti...

Страница 2: ...s 5 3 Electrical Requirements 5 4 Unpacking the System 5 5 Power Cord Power Connection 5 6 Foot Control Cable Assembly Connection 5 7 Water Supply Line Connection 6 SPS Ultrasonic Scaler Description 6...

Страница 3: ...rocedures 2 2 Warnings Persons fitted with cardiac pacemakers have been cautioned that some types of electronic equipment might interfere with the operation of a pacemaker Although no instance of pace...

Страница 4: ...ther areas contact your local DENTSPLY representative 4 2 GeneralInfectionControl Recommendations As with all dental procedures use standard personal protection equipment i e wear face mask eyewear or...

Страница 5: ...into an approved outlet Section 5 Installation Instructions cont d 5 6 Foot ControlCable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in un...

Страница 6: ...o position Foot Control provides the capability to increase the front panel Power dial setting When lighted indicates Foot Control is fully depressed and System is functioning in Boost mode Lavage Flo...

Страница 7: ...st motion DO NOT FORCE 6 3 Cavitron 30K Ultrasonic Inserts The many styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications See enclosed booklet fo...

Страница 8: ...depressed so adequate fluid will be available to cool tip tooth interface In general it is suggested that a feather light touch be used both supra and subgingivally The motion of the activated tip an...

Страница 9: ...lying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufacturer To clean system generously spray disinfectant solutio...

Страница 10: ...epth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for operating theatres Protection Class 1PXO Operating Environ...

Отзывы: