Section 1: Indications

Section 2: Contraindications and Warnings

Section 3: Precautions

3.1 Precautions for All Systems

• Do not place the System on or next to a radiator or

other heat source. Excessive heat may damage the

System’s electronics. Place the System where air is

free to circulate on all sides and beneath it.

• The System is portable, but must be handled

with care when moving.

• Equipment flushing and dental water supply system

maintenance are strongly recommended. See Water

Supply under Section 9: System Care

• Close the coolant water shut-off valve to the System

every night before leaving the office. The use of an

in-line water filter is recommended.

2.1 Contraindications

• Ultrasonic Systems should not be used for restorative

dental procedures involving the condensation of

amalgam.

• For optimum performance use only inserts

manufactured by DENTSPLY

®

 Professional.

Ultrasonic procedures:

• All general supra and subgingival scaling applications.
• Periodontal debridement for all types of periodontal diseases.
• For patients with a history of sensitivity to ultrasonics.
• Endodontic procedures.

2.2 Warnings

• Persons fitted with cardiac pacemakers have been

cautioned that some types of electronic equipment

might interfere with the operation of a pacemaker.

Although no instance of pacemaker interference has

ever been reported to DENTSPLY

®

, we recommend

that the handpiece and cables be kept at least 6 to 9

inches (15 to 23 cm) away from any pacemaker and

pacemaker leads during use.

• There are a variety of pacemakers on the market.

Clinicians should contact the pacemaker manufacturer

or the patient's physician for detailed information about

the pacemaker.

3.2 Precautions for Ultrasonic

Prophylaxis Procedures

• Like a toothbrush, ultrasonic inserts “wear out” with

use. Inserts with just 2 mm of wear lose about 50% of

their scaling efficiency. In general, it is recommended that

ultrasonic inserts be discarded and replaced after 2 mm

of wear to maintain optimal efficiency and avoid

breakage.

• If excessive wear is noted, or the insert has been bent,

reshaped or otherwise damaged, discard the insert

immediately.

• Ultrasonic insert tips that have been bent, damaged, or

reshaped are susceptible to in-use breakage and should

be discarded and replaced immediately.

• Retract the lips, cheeks and tongue to prevent contact

with the insert tip whenever it  is placed in the patient’s

mouth.

Содержание Cavitron SPS

Страница 1: ...ically controls the tip stroke over different working conditions The Cavitron SPS scaler produces 30 000 microscopically small strokes per second at the insert s working tip This combined with acousti...

Страница 2: ...s 5 3 Electrical Requirements 5 4 Unpacking the System 5 5 Power Cord Power Connection 5 6 Foot Control Cable Assembly Connection 5 7 Water Supply Line Connection 6 SPS Ultrasonic Scaler Description 6...

Страница 3: ...rocedures 2 2 Warnings Persons fitted with cardiac pacemakers have been cautioned that some types of electronic equipment might interfere with the operation of a pacemaker Although no instance of pace...

Страница 4: ...ther areas contact your local DENTSPLY representative 4 2 GeneralInfectionControl Recommendations As with all dental procedures use standard personal protection equipment i e wear face mask eyewear or...

Страница 5: ...into an approved outlet Section 5 Installation Instructions cont d 5 6 Foot ControlCable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in un...

Страница 6: ...o position Foot Control provides the capability to increase the front panel Power dial setting When lighted indicates Foot Control is fully depressed and System is functioning in Boost mode Lavage Flo...

Страница 7: ...st motion DO NOT FORCE 6 3 Cavitron 30K Ultrasonic Inserts The many styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications See enclosed booklet fo...

Страница 8: ...depressed so adequate fluid will be available to cool tip tooth interface In general it is suggested that a feather light touch be used both supra and subgingivally The motion of the activated tip an...

Страница 9: ...lying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufacturer To clean system generously spray disinfectant solutio...

Страница 10: ...epth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for operating theatres Protection Class 1PXO Operating Environ...

Отзывы: