ENGLISH • 5
5.3 Air Line Requirements &
Recommendations
• An air supply line is factory installed on Cavitron
™
JET
Combination System. Do not disconnect it.
• Incoming air supply line pressure to the System must be
65 psig (448kPa) to 100 psig (690kPa). If your office air
line pressure is above 100 psig (690kPa), install an air
pressure regulator on the supply line to your Cavitron
™
JET Combination System.
• A manual shut-off valve on the office air supply line
should be used so that the air line can be completely
shut-off, and the line pressure relieved when the office is
unoccupied.
• Your System must be supplied with clean, dry air. To
help prevent water condensation from forming in the air
supply line which may cause the system to malfunction, it
is strongly recommended that an air dryer be used on the
compressor line supplying the System.
5.4 Electrical Requirements
Refer to Section 11: Specifications.
5.7 Foot Control Cable Assembly
Connection
Align the Foot Control plug with the receptacle on the back of
the System and push it in until firmly seated.
5.5 Unpacking the System
Carefully unpack your Cavitron
™
JET Combination System and
verify that all components and accessories are included:
1. Cavitron
™
JET Combination System with water
supply line, air supply line, handpiece and cable
connector assembly factory installed.
2. Air filter assembly.
3. Detachable AC Power Cord set.
4. Cavitron
™
JET Insert Nozzle with cleaning tool.
5. Dual position Foot Control Assembly.
6. Cavitron
®
Ultrasonic Inserts (quantity optional).
7. Efficiency Indicator for Cavitron
®
Inserts.
8. JetShield™ Aerosol Reduction Device.
9. Literature Packet.
ON/OFF rocker switch
Section 5: Installation Instructions cont'd
5.6 Power Cord/Power Connection
• Verify the Power Control ON/OFF rocker switch located
at the left front underside of the System is set to the OFF
position before proceeding.
• Plug the detachable AC cord into the back of the System.
• Plug the 3 prong grounded plug into an approved outlet.