background image

ENGLISH  • 7

Section 6: Cavitron

-Jet Combination System Ultrasonic Scaler

and Air Polishing Prophylaxis System cont'd

O-Ring

Provides seal for handpiece

coolant.

Insert Tip

Shape and size of tip determines access and

adaptation. Preheated Lavage directed to

tip.

Finger Grip

Connecting Body

Transfers and amplifies mechanical

motion of stack to insert tip.

6.3 Cavitron

®

 

30K

 Ultrasonic Inserts

The many styles of Cavitron

®

 30K

 Ultrasonic Inserts are easily interchangeable

for various procedures and applications. See enclosed booklet for specific information.

Hold the handpiece in an upright position. Activate to bleed any air bubbles that might be trapped inside the handpiece. Lubricate the
rubber O-ring on the insert with water before placing it into the handpiece.  Fully seat insert with a gentle push-twist motion.  DO NOT
FORCE.

6.4 Cavitron

 JET Air Polishing Insert nozzle

The Air Polishing Insert Nozzle automatically selects the JET mode when placed in handpiece.

Cleaning Powder Delivery Tube: Directs

air/powder flow to insert tip.

Insert Heater Rod: Heats the

delivered water for patient

comfort.

O-Ring: Seals water when Insert Nozzle is

fully seated in Handpiece.

Air Polishing Insert Tip:

Tube-in-a-tube design

deliveres precise

air/water/powder mixture

at point of delivery.

6.2 Handpiece

The handpiece accepts all Cavitron

®

 30K

 Ultrasonic Inserts and the JET air polishing insert nozzle. The open end of the handpiece

accepts all Cavitron

®

 30K

 inserts.

Insert Port

Air Polishing Powder Delivery Port

Powder Flow Window

Handpiece

Cable

Magnetostrictive Stack

Converts energy provided by the hand-

piece into mechanical oscillations

used to activate the insert tip.

Insert Marking

Manufacturer, Date (YYMM),

Frequency, Type.

E.g. DENTSPLY 9608 30K FSI-SLI-10S

Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Place the heater rod into the insert port of the
handpiece. Align the powder delivery tube with the powder delivery port and gently push into handpiece until fully seated.

Содержание Cavitron JET

Страница 1: ...oigneusement et compl tement avant d utiliser le syst me Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad Syst me combin Cavitron JET D tartreur ultras...

Страница 2: ...e insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional oper...

Страница 3: ...nection 6 Cavitron JET Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System 6 8 6 1 System Controls 6 2 Handpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishin...

Страница 4: ...Do not direct the air polishing stream at soft tissue or into the sulcus Tissue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulc...

Страница 5: ...ystem s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patien...

Страница 6: ...Specifications 5 7 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated 5 5 Unpacking the System Carefull...

Страница 7: ...upply line to office air supply line or a Cavitron DualSelect dispensing System Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaleran...

Страница 8: ...e automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs air powder flow to insert tip Insert Heater Rod Heats the delivered water for patient comfort O Ring...

Страница 9: ...perationforUltrasonicProphylaxisProcedures No ultrasonic energy or lavage flow Ultrasonics and lavage flow are activated Foot control depressed all the way 6 7 Operation for Air Polishing Procedures F...

Страница 10: ...he rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE Set the Unit s Power Adjustment knob to the lowest power s...

Страница 11: ...ing of the powder chamber An additional cap without a hole is provided for storage It is strongly recommended that the Powder Chamber be emptied at the end of each day This will reduce moisture absorp...

Страница 12: ...ly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes re...

Страница 13: ...t s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Use a saliva ejector or HVE In the posterior regions use the cheek to help contain aerosol spray 8 7 Perf...

Страница 14: ...ectant solution to air dry 5 If using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut Down Procedures at the end of the day 1 Turn the Unit OFF 2 Rem...

Страница 15: ...in the Foot Control Receptacle on the back of the System System operates No lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s de...

Страница 16: ...Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for ope...

Страница 17: ...6 Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Ph...

Отзывы: