Section 4: Infection Control

4.1 Infection Control Information

Reference Card

For your convenience, an Infection Control Information reference
card, printed on durable stock, has been included with your
Cavitron

 JET Combination System. If your card is missing or

misplaced, you can request a replacement in the U.S. by calling
Customer Service at 1-800-989-8826 Monday through Friday,
8:00 AM to 5:00 PM (Eastern Time) For other areas, contact
your local DENTSPLY representative.

4.2 General Infection Control

Recommendations

• As with all dental procedures, use standard personal

protection equipment (i.e., wear face mask, eyewear or face
shield, gloves and protective gown).

4.3 Water Supply Recommendations

It is highly recommended that all dental water supply systems
conform to applicable CDC (U.S. Centers for Disease Control
and Prevention) and ADA (American Dental Association)
standards, and that all recommendations be followed in terms of
flushing, chemical flushing, and general infection control
procedures. See sections 4.2 and 9.

Section 3: Precautions cont'd

3.3 Precautions for Air Polishing Procedures

• Only set the air polishing powder flow control to the

maximum position (H) when it is necessary to remove
particularly difficult stains. Return the powder flow
control to the medium position upon the completion of
the procedure.

• Avoid prolonged or excessive use on restorative dental

materials.

• As with high speed handpieces, and other dental devices,

the combination of water and ultrasonic vibration from
your Cavitron

 JET Combination System will create

aerosols. With proper technique, much of the Cavitron

JET Combination System’s aerosol dispersion can be
effectively controlled and minimized. Please carefully
follow the procedural guidelines in this manual regarding
the use of your System.

• For maximal operator and patient safety, carefully follow

the Infection Control Information procedures detailed
on the reference card accompanying your System.

• Avoid prolonged use on cementum or dentin.
• Avoid prolonged use on polished metal surfaces as a

matte finish may result.

• Patients who are on a restricted sodium diet, or have

severe respiratory illness, should consult their physician
before undergoing air polishing prophylaxis procedures.

• Patients wearing contact lenses should remove them

prior to air polishing treatment.

Section 5: Installation Instructions

5.1 General Information

If the installation of your Cavitron

 JET Combination System is

performed by someone other than trained DENTSPLY

®

 Dealer

personnel, care should be taken to observe the following require-
ments and recommendations.

5.2 Water Line Requirements

• The System’s water supply line is factory installed. Do

not disconnect from the System.

• Incoming water supply line pressure to the System must

be 25 psi (172 kPa) to 60 psi (414kPa). If your dental
water system’s supply line pressure is above 60 psi
(414 kPa), install a water pressure regulator on the supply
line to your Cavitron

 JET Combination System.

• A manual shut-off valve on the dental water system

supply  line should be used so that the water can be
completely shut-off when the office is unoccupied.

• A filter in the dental water system supply line is recom-

mended so that any particulates in the water supply will
be trapped before reaching the System.

• After the above installations are completed on the dental

water supply system, the dental office water line should
be thoroughly flushed prior to connection to the System.

Содержание Cavitron JET

Страница 1: ...oigneusement et compl tement avant d utiliser le syst me Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad Syst me combin Cavitron JET D tartreur ultras...

Страница 2: ...e insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional oper...

Страница 3: ...nection 6 Cavitron JET Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System 6 8 6 1 System Controls 6 2 Handpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishin...

Страница 4: ...Do not direct the air polishing stream at soft tissue or into the sulcus Tissue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulc...

Страница 5: ...ystem s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patien...

Страница 6: ...Specifications 5 7 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated 5 5 Unpacking the System Carefull...

Страница 7: ...upply line to office air supply line or a Cavitron DualSelect dispensing System Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaleran...

Страница 8: ...e automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs air powder flow to insert tip Insert Heater Rod Heats the delivered water for patient comfort O Ring...

Страница 9: ...perationforUltrasonicProphylaxisProcedures No ultrasonic energy or lavage flow Ultrasonics and lavage flow are activated Foot control depressed all the way 6 7 Operation for Air Polishing Procedures F...

Страница 10: ...he rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE Set the Unit s Power Adjustment knob to the lowest power s...

Страница 11: ...ing of the powder chamber An additional cap without a hole is provided for storage It is strongly recommended that the Powder Chamber be emptied at the end of each day This will reduce moisture absorp...

Страница 12: ...ly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes re...

Страница 13: ...t s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Use a saliva ejector or HVE In the posterior regions use the cheek to help contain aerosol spray 8 7 Perf...

Страница 14: ...ectant solution to air dry 5 If using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut Down Procedures at the end of the day 1 Turn the Unit OFF 2 Rem...

Страница 15: ...in the Foot Control Receptacle on the back of the System System operates No lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s de...

Страница 16: ...Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for ope...

Страница 17: ...6 Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Ph...

Отзывы: