ENGLISH • 6
6.1 System Controls
Top Grooves
Provide a convenient, temporary place-
ment for the handpiece during treatment
procedures. Always place handpiece in
holder when not being used.
Power Adjustment
Turn knob to select ultrasonic
power level for operation:
clockwise increases system
power, counter clockwise
decreases system power.
NOTE: For Air Polishing, set
the power level between
medium (mid) and maximum.
Handpiece
Holds all Cavitron
®
30K™
Ultrasonic Inserts and transmits
power and lavage from the
System to the insert. It also holds
the Cavitron™ JET insert nozzle
which automatically selects air
polishing mode.
Section 5: Installation Instructions cont'd
5.8 Water Supply Line Connection
• Connect the free end of the system’s water supply line to
office water supply line or a Cavitron
®
DualSelect™
Dispensing System.
• Inspect all connections to make certain there are no leaks.
5.9 Air Supply Line Connection
• Connect the free end of the system’s air supply line to
office air supply line or a Cavitron™ DualSelect
™
dispensing System.
• Inspect all connections to make certain there are no leaks.
Section 6: Cavitron
™
Jet Combination System Ultrasonic
Scaler and Air Polishing Prophylaxis System
Description
AC Power Indicator Light
Lighted when System is
switched ON.
Power Control
ON/OFF switch located at the left
front underside of the System.
Boost Indicator Light
The two-position Foot Control provides the
capability to increase the front panel Power
dial setting. When lighted, indicates Foot
Control is fully depressed, and System is
functioning in "Boost" mode.
Lavage Flow Adjustment
Turn knob to select Lavage Flow level during System operation.
Clockwise increases flow at insert tip, counter clockwise decreases flow.
The rate of flow through the Handpiece determines the temperature of the
lavage. Low flow rates produce warmer temperatures, higher flow rates
produce cooler temperatures.
Powder Fill Cap
Removable cap for filling and emptying the
powder chamber. An adjustable pointer
increases or decreases powder flow rates.
Handpiece Holder
Securely holds the System's
Handpiece when system is
not being used.