5 Care, cleaning and maintenance by the user
5.1 Basics
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
74
66 82 384 D3684
D3684.201.01.02.02 07.2019
Daily
Purge the water lines (auto-purge function, purge function) or purge the
water lines manually
Flush water lines
Clean the suction system using an external tank
Clean the suction hoses
Disinfect the dentist element
weekly
Change the cotton roll on the high-speed handpiece hose and oil
collector
Monthly or as required
Change the water and air filters
Microbiological water test
5.1.2
Care, cleaning, and disinfecting agents
Unsuitable care and cleaning agents or disinfectants may corrode the
surface of the unit or impair its functioning.
Please therefore only use care, cleaning, and disinfecting agents
approved by the manufacturer. For more information, please refer to the
section "Care, cleaning agents, and disinfectants" [→ 97].
NOTE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by
Dentsply Sirona!
A continuously updated list of approved products can be obtained via
the Internet at
"www.sirona.com". In the navigation bar, follow the menu
path
"SERVICE" / "Technical Documentation" / "General documents"
and then open the document
"Care, cleaning and disinfection agents".
If you do not have access to the Internet, please contact your customer
service center (see Sect. 1.2) to order the list (REF 59 70 905).